Vital Signs Monitor User’s Manual
Chapter 7-5
1. Pressure unit
2. Systolic blood pressure
3. Mean arterial blood pressure
4. Pulse Rate (obtained from NIBP)
5. Diastolic blood pressure
7.7 Setting NIBP
Select NIBP to enterNIBP Setupmenu.
7.7.1 Setting Unit
InNIBP Setupmenu, setUnittommHgorkPa.
7.7.2 Setting Initial Press
In
NIBP Setup
menu, set
Init Press
to a desired value.
7.8 Setting Venipuncture Press
You can use the NIBP cuff to cause sub-diastolic pressure, and block the venous
blood vessel to assist venous puncture.
1. Select “NIBP to enterNIBP Setupmenu.
2. Set
Veni. Press
to a desired value.
3. SelectVenipuncture.
4. Puncture vein and draw blood sample.
5. Select
Venipuncture
again to deflate the cuff. The cuff deflates
automatically after a set time if you do not deflate it.
7.9 NIBP Resetting
Select Reset inNIBP Setupmenu to restore the inflation value of blood
pressure pump to the initial value. In case the blood pressure pump doesn’t work as
normal but without any prompt, the blood pressure pump can be checked by reset,