Vital Signs Monitor User’s Manual
Chapter 7-5
2. Systolic blood pressure
3. Mean arterial blood pressure
4. Pulse Rate (obtained from NIBP)
5. Diastolic blood pressure
Select ”NIBP” to enter【NIBP Setup】menu.
7.7.1 Setting Unit
In【NIBP Setup】menu, set【Unit】to【mmHg】or【kPa】.
7.7.2 Setting Initial Press
In
【
NIBP Setup
】
menu, set
【
Init Press
】
to a desired value.
7.8 Setting Venipuncture Press
You can use the NIBP cuff to cause sub-diastolic pressure, and block the venous
blood vessel to assist venous puncture.
1. Select “NIBP” to enter【NIBP Setup】menu.
2. Set
【
Veni. Press
】
to a desired value.
4. Puncture vein and draw blood sample.
5. Select
【
Venipuncture
】
again to deflate the cuff. The cuff deflates
automatically after a set time if you do not deflate it.
Select【 Reset 】in【NIBP Setup】menu to restore the inflation value of blood
pressure pump to the initial value. In case the blood pressure pump doesn’t work as
normal but without any prompt, the blood pressure pump can be checked by reset,