V6 Vital Signs Monitor Users Manual
7
Measurement range
0 g/dl to 25 g/dl
Accuracy
8 g/dl to 17 g/dl±1%non-motion conditions
8 g/dl or 17 g/dlunspecified
SpOC
Measurement range
0 ml/dl to 35 ml/dl
A.4.2 NIBP
Measurement way
Automatic oscillometry
Measurement range
Adult
Sys
(30270) mmHg
Dia
(10220) mmHg
Map
(20235) mmHg
Pediatric
Sys
(30235) mmHg
Dia
(10220) mmHg
Map
(20225) mmHg
Neonatal
Sys
(30135) mmHg
Dia
(10110) mmHg
Map
(20125) mmHg
Cuff pressure range
(0300) mmHg
Resolution
1 mmHg
Pressure accuracy
Static: ±3 mmHg
Clinic: Average error: ±5 mmHg, standard deviation: ≤8 mmHg
Unit
mmHg, kPa
PR range
(40240) bpm
PR Resolution
1bpm
Cuff auto deflation
The cuff will deflate automatically when power is off or time of
measurement is beyond 120 seconds (90 seconds for neonate) or the
cuff pressure is beyond the overpressure protection set by software
and hardware.
Measurement time
Normally, it is 20s to 45s (depending on HR and moving
interference typically)
Overpressure protection
Adult
Pediatric
Neonatal
(297±3) mmHg
(252±3) mmHg
(147±3) mmHg