Vital Signs Monitor User’s Manual
Chapter 3-6
——An alarm condition destined occurs.
——The monitor is not in the state of alarm paused or system silence.
The nurse call function should not be used as the primary patient alarm
inform source. It is necessary for combining the auditory and visual
alarm signal and the patient clinical feature and symptom as the primary
information to medical and nursing staff about the physiological
condition of the patient.
3.8.1 Turning on/off Nurse Call
1. Select【Menu】→【System Setup】→【Machine Mainte】→enter the required
2. Select【Factory Mainte】→enter the required password.
3. Select【Nurse Call】, set it to 【on】.
3.8.2 Setting Nursecall Type
1. Select【Menu】→【System Setup】→【Machine Mainte】→enter the required
2. Select【Nurse Call】to enter the nurse call menu.
3. Set 【Nursecall Type】to 【Normally Open】or 【Normally Closed】.
3.8.3 Setting Call Time
1. Select【Menu】→【System Setup】→【Machine Mainte】→enter the required
2. Select【Nursecall Setup】to enter nurse call setup menu.
3. Set【Call Time】to【1 Sec】or【Continuous】.
3.8.4 Triggering Nurse Call
1. Select【Menu】→【System Setup】to enter system setup menu.
2. Select【Nursecall Setup】to enter nurse call setup menu.
3. Set【PhyAlarm Trigger】and【TecAlarm Trigger】to【Off】, 【Low】, 【Med】