V6 Vital Signs Monitor Users Manual
11
B.4.2 Special Setups (Masimo)
SpO
2
setup
Factory defaults
Sensitivity Mode
Normal
Fast Sat Mode
off
Smart Tone Mode
off
Waveform Mode
Resp.out
Alarm Delay
off
SpHb Mode
Arterial
SpHb Average Time
Long
SpHb Precision
0.1
SpHb Unit
g/dL
B.5 NIBP
NIBP setup
Adult
Pediatric
Neonatal
Alarm switch
On
Alarm Level
Medium
Alarm Print
Off
High alarm limit of Sys
160 mmHg
120 mmHg
90 mmHg
Low alarm limit of Sys
90 mmHg
70 mmHg
40 mmHg
High alarm limit of Map
110 mmHg
90 mmHg
70 mmHg
Low alarm limit of Map
60 mmHg
50 mmHg
25 mmHg
High alarm limit of Dia
90 mmHg
70 mmHg
60 mmHg
Low alarm limit of Dia
50 mmHg
40 mmHg
20 mmHg
Measure Mode
Manual
Unit
mmHg
Interval
15 min
Color
red
Venipuncture Press
60 mmHg
40 mmHg
30 mmHg
Inflation
170 mmHg
130 mmHg
100 mmHg