Communication Error Codes | 85
Code Description
than carrier clear
height
2C41 Mixed Store and
Park function - lift
height less than
carrier clear
height
* Does not apply to Stay-on Lid mode.
These are the basecode limits for the plate geometry parameters. The BioStack's
Specifications are typically tighter to prevent errors:
MAX_HEIGHT 24.00 mm
MIN_HEIGHT 5.00 mm
MAX_WIDTH 87.0 mm
MIN_WIDTH 84.30 mm
MAX_GRIP_WIDTH 86.30 mm
MIN_GRIP_WIDTH 80.80 mm
MIN_SKIRT 1.00 mm
Washer or Dispenser Communication Error Codes
The error codes listed below may appear on the display of washers and dispensers
during operation with the BioStack.
n On instruments with keypads: If a four-digit code is prefixed with B- or
“BIOSTACK ERROR” appears, the code was generated by the BioStack.
Otherwise, the code was most likely generated by the interfacing instrument;
refer to that instrument’s operator’s manual.
Code Probable Cause
8300 An invalid password was sent to the BioStack during a protected operation. This
error can only occur when the BioStack is under serial control, using a software
program other than the BioStack PC Control Software.
8301 The BioStack was requested to send some configuration data to the controlling
software, and the BioStack did not know about the requested data.
8302 A checksum was generated following a code download, which did not match what
BioTek Instruments, Inc.