EasyManua.ls Logo

Biotek BioStack 3 - Plate De-Lidding: How It Works

Biotek BioStack 3
127 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
50 | Chapter 3: Operation
the plate, which could lead to false results. Review these guidelines: Prevent
Condensation on Plate Lids on the facing page.
Plate De-Lidding: How It Works
BioStack 4 Only.
Here is an overview of how the BioStack 4 handles plates with lids. Like all
BioStack processing, plates are pulled from the bottom of the input stack, i.e., the
bottom plate is the first one transferred, and returned to the bottom of the output
stack.
Important: Beware of condensation forming on lids. Do not process plates with lids when
the application causes condensation on the lids: the condensate may drip and contaminate
the plate, which could lead to false results. Review these guidelines: Prevent
Condensation on Plate Lids on the facing page.
n When processing plates with lids the BioStack 4 performs several extra error-
checking movements, using its fingers to determine if the expected labware,
i.e., a plate, lid, or combination, is present.
Cross section of microplates with lids in the input stack
BioStack™ Microplate Stacker

Table of Contents

Related product manuals