82 | Chapter 6: Troubleshooting & Error Codes
Code Probable Causes
2900 Invalid combination error. An invalid combination of parameters associated
with carrier trays and stacks was attempted. An internal code error is likely.
Contact BioTek TAC.
2A0# Dip switch SW1-4 is in the wrong position. See Dip Switch Settings on
page 28.
2C## Plate Lid Definition exceeds limits. Applies only when processing plates with
lids. See Plate Lid Definition Errors below.
A100/AE#0 Task has been deleted because of an error; fatal. Internal code malfunction.
Contact BioTek TAC.
A30# Fatal error. Internal code malfunction. The given motor number was not
available. # = motor; see page 87 for a list of motor numbers. Contact
BioTek TAC.
AE8# The motor profile resolution is invalid for phase correction. # = motor; see
page 87 for a list of motor numbers. Contact BioTek TAC.
Plate Lid Definition Errors
When processing microplates with lids, correct the input values for the offending parameter in
the Plate Lid Definition file when one of these error codes is displayed. See the legend below
for referenced values.
Code Description
2C01 Plate Height <
MIN_HEIGHT or
> MAX_HEIGHT
2C02 Long Side Skirt
Height >= Plate
Height or < MIN_
SKIRT
2C03 Short Side Skirt
Height >= Plate
Height or < MIN_
SKIRT
2C04 Lid Grip Width <
MIN_GRIP_
WIDTH or >
MAX_GRIP_
WIDTH
BioStack™ Microplate Stacker

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Biotek BioStack 3 and is the answer not in the manual?

Biotek BioStack 3 Specifications

General IconGeneral
BrandBiotek
ModelBioStack 3
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals