Communication Error Codes | 81
Code Probable Causes
2300 Gripper width saved is outside allowed range. Incorrect measurement
entered. Incorrect units.
2401 Hardware not present. A dual carrier is not present. An internal code error is
likely. Contact BioTek TAC.
2402 Hardware not present. An odd spaced dual carrier is not present. An internal
code error is likely. Contact BioTek TAC.
2403 Hardware not present. A wrist is not installed. An internal code error is likely.
Contact BioTek TAC.
2404 A barcode reader is not present. See Dip Switch Settings on page 28 and
make sure the setting is correct for the barcode scanner. If problems persist,
contact BioTek TAC.
2405 Alternative wrist position is not selected. See Wrist Angle Setting
Requirements on page 26 and make sure the dip switch setting for wrist
angle matches your instrument requirement. Contact BioTek TAC if problems
persist.
2600 A gripper interface dimension is out-of-range at power up. Rerun the
Instrument Alignment Utility to clear the error.
2601 The instrument’s gripper release position is invalid. Contact BioTek TAC.
2602 The instrument’s gripper clear position is invalid. Contact BioTek TAC.
2603 The instrument’s gripper grip position is invalid. Contact BioTek TAC.
2700 Error attempting to run the barcode scanner SET command. A response
returned from the scanner is invalid.
2701 A response from the barcode scanner while attempting the SET command
does not contain a valid barcode location. Valid numbers are 03.
2702 The barcode type returned by the scanner is not one of the four types
supported by the scanner. See Preparing and Attaching Barcode Labels
on page 99 for required barcode types.
2703 Barcode type is not supported. One of the four default barcode types is not
supported by the scanner. See Preparing and Attaching Barcode Labels
on page 99 for required barcode types.
2704 Error disabling start/stop character transmission. While attempting to tell the
scanner to disable the transmission of start and stop characters along with
the barcode value for the Codeabar barcode type, an error occurred.
2800 Out of range error. An attempt was made to set a configuration parameter,
using an out-of-range value.
BioTek Instruments, Inc.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Biotek BioStack 3 and is the answer not in the manual?

Biotek BioStack 3 Specifications

General IconGeneral
BrandBiotek
ModelBioStack 3
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals