140 | Chapter 6: Troubleshooting and Error Codes
BioTek Instruments, Inc.
General Errors
General errors indicate nonfatal conditions that require attention.
Code Description and Probable Causes
0100 Abort
The read or task has been aborted.
0101 Abort Error
The read or task has been aborted. The 0101 indicates a software abort.
Probable Causes:
User aborted read from KC4.
User aborted from another serial interface.
0200 X-axis (carrier in/out) motor did not find the opto (home) sensor
A motor was unable to move to its “home” position as registered by feedback
from an optical sensor.
Probable Causes:
Defective or broken optical sensor.
Flex cable is disconnected or damaged.
X-axis movement is limited so that the optical sensor cannot be
interrupted.
Note: In cases where a sensor is not functioning, the
motor will drive the axis to its mechanical stop and
generate substantial noise.

Table of Contents