Define | 45
ELx800 Operator’s Manual
Programming Note
Assays with certain criteria may “lock up” during the “Calculating Results…”
phase of data reduction (see
Beginning the Plate Read on page 90). This can be
corrected by modifying assay criteria as follows:
Specify
at least one (01) sample when mapping the plate, and/or
Change PROMPT SAMPLE COUNT? to YES.
Set CURVE FIT to NONE if you do not have a standard curve as part of the
assay.
1 From the Main Menu, press DEFINE MAP, and continue to press
ENTER until ENTER NUMBER OF SAMPLES is displayed. Enter at least
one sample.
The software will lock up if zero (00) samples are selected.
E N T E R N UMBER OF
S A M P L E S : 0 1
2 From the Main Menu, press UTIL READ, and continue to press
ENTER until PROMPT SAMPLE COUNT? is displayed. Select YES.
P R O M P T SAMPLE COUNT? Y E S
Y E S NO
3 From the Main Menu, press DEFINE SELECT ASSAY NUMBER:
ENTER. At the DEFINE: menu, press CURVE to display CURVE-FIT
TYPE: and select NONE.
C U R V E - F IT TYPE: NONE
N O N E L INEAR QUAD *M O R E