6 | Chapter 1: Introduction
BioTek Instruments, Inc.
Specifications
Standard Model
Wavelength Range:
400 to 750 nm
Filters:
10 nm half-bandwidth interference filters.
User-accessible filter wheel. Up to 5 filters may be
installed on the instrument at one time.
Filters supplied: 405 nm, 450 nm, 490 nm and 630 nm.
• The following specifications apply to 96-well, flat- or round-bottom plates,
single-wavelength measurements with a 50-second read (normal read
mode).
Absorbance Measurement
Range:
0.000 to 3.000 OD
Accuracy: ± 1.0% ± 0.010 OD from 0.000 to 2.000 OD
@ 405 nm
Linearity: ± 1.0% from 0.000 to 2.000 OD at 405 nm
± 3.0% from 2.000 OD to 3.000 OD @ 450 nm
Repeatability (STD): ± 0.5% ± 0.005 OD from 0.000 to 2.000 OD @ 405 nm
• The following specifications apply to 96-well, flat- or round-bottom plates,
single-wavelength measurements with a 30-second read (rapid read mode).
Absorbance Measurement
Range:
0.000 to 3.000 OD
Accuracy: ± 2.0% ± 0.020 OD from 0.000 to 2.000 OD @ 405 nm
Linearity: ± 2.0% from 0.000 to 2.000 OD @ 405 nm
Repeatability (STD): ± 1.0% ± 0.010 OD from 0.000 to 2.000 OD @ 405 nm
• Light Source: Tungsten gas-filled bulb
• Dimensions: 42 cm x 38 cm x 18 cm (16.5" deep x 15" wide x 7" tall)
• Weight: 8 kg (18.5 lb. Maximum)
• Environment: Operating temperature 18° to 40°C
• Humidity: 10% to 85% noncondensing
• Power Supply: Input 100 to 240 V~ ± 10.0% @ 50 to 60 Hz
Output +24 VDC, 2.1 A