26 | Chapter 2: Installation
BioTek Instruments, Inc.
Checking the Reader’s Filter Table Setting
After installing new filters, ensure that the ELx800’s filter table (the reader’s
software reference for filter locations) matches the physical location of the filters.
To check or change the software filter table:
1 Power up the reader. At the Main Menu Screen, press UTIL to display
the SELECT UTILITY OPTION menu.
R E A D Y 9:45AM 01/31 / 0 3
R E A D D EFIN
E
REPORT U T I L
S E L E C T UTILITY OPTION :
T E S T S S ETUP OUTPUT RE A D
2 From the SELECT UTILITY OPTION menu, press SETUP. The Edit
Setup Information screen appears on the display.
E D I T S E TUP INFORMATIO N :
D A T E TIME FILTER *M O R E
3 From this menu, press FILTER.
4 The wavelength for Filter #1 will be displayed. To advance to Filter 2,
press ENTER.
5 To change the filter wavelength number, use the NUMERIC keypad to
enter a number at the cursor location. The cursor will automatically
advance to the next editable field. Press ENTER to save the entry and
move to the next filter on the filter table.