Specifications | 9
ELx800 Operator's Manual
The following specifications apply to a 384-well plate. The minimum volume of
fluid in a well to obtain the specified performance is 80 microliters. The
specifications apply to the dual wavelength mode of read only.
Absorbance Measurement
Range:
0.000 to 3.000 OD
Normal Read Mode:
Dual wavelength (4 minutes, 45 seconds)
Accuracy: ± 2.0% ± 0.020 OD from 0.000 to 2.000 OD
@ 405 nm
Linearity: ± 2.5% from 0.000 to 2.000 OD @ 405 nm
Repeatability (STD): ± 1.5% ± 0.010 OD from 0.000 to 2.000 OD
@ 405 nm
Rapid Read Mode:
Dual wavelength (3 minutes, 25 seconds)
Accuracy: ± 2.5% ± 0.020 OD from 0.000 to 2.000 OD
@ 405 nm
Linearity: ± 2.5% from 0.000 to 2.000 OD @ 405 nm
Repeatability (STD): ± 2.0% ± 0.010 OD from 0.000 to 2.000 OD
@ 405 nm