102 | Gravimetric Calibration Instructions
MicroFlo Select Microplate Dispenser
Figure 2: Measure and record the weight of each of the strips prior to dispensing.
Tip: Label the strips with a letter for each row, 1 – 8, to make it easier to keep track of
them when weighing each one individually.
5 Place each strip on the scale and record its mass (using the provided
worksheets). This will be the
Tare mass for each of the strips.
6 Put the tube organizer into the test fluid supply vessel.
7 Prime the tubing until any visible air bubbles are removed.
8 Run the dispense protocol.
Figure 3: Measure and record the weight of each of the strips after dispensing.