18 System User Guide Rev. 2.5 Kolibri Ver. 2.0
System Components & Correct Handling
Restrictions to
Environment
System components are not suitable for use in the presence of flammable anesthetic
mixtures containing air, oxygen, nitrous oxide or laughing gas.
Do not let liquids entering the system components.
Do not place system components on an unstable cart, stand, or table. The system
might fall down and be seriously damaged.
Interferences
The system produces electromagnetic fields that may interfere with other sensitive
equipment, and can itself be disturbed by other electromagnetic fields.
Due to EMC (electromagnetic compatibility) reasons, do not use system components
adjacent to or stacked with other equipment. If it is unavoidable to stack the system
or use it adjacent to other equipment, verify that the workstation operates normally.
The camera’s infrared light may interfere with other IR-based OR equipment, such as
remote controls, pulse oximeters, IR sensitive microscopes.
If other devices are interfered by the camera’s infrared light, reposition these devices and/or the
camera so that interference effects are no longer visible.