System User Guide Rev. 2.5 Kolibri Ver. 2.0 31
KOLIBRI WORKSTATION
3.3.5 Restrictions on Connecting Equipment to Panels
Restrictions
Additional equipment connected to medical electrical equipment must comply with
the respective IEC or ISO standards (e.g., IEC 60950 for data processing equipment
and IEC 60601-1 for medical equipment). Furthermore, all configurations shall comply
with the requirements for medical electrical systems (see IEC 60601-1-1 or clause 16
of the 3rd edition of IEC 60601-1, respectively). Anyone connecting additional equip-
ment to medical electrical equipment configures a medical system and is therefore re-
sponsible that the system complies with the requirements for medical electrical
systems. Please note that local laws take priority over the above mentioned require-
ments. In doubt, consult your local representative or the technical service department.
NOTE: Unless otherwise specified, the use of multiple socket outlet(s) is not permitted.
Safe Handling of
Cables
Unplug and wind up all cables when system is not in use or before moving it. Do not
pull cables.
USB
Only connect passive USB devices to the system, otherwise safety and effectiveness
of the equipment cannot be guaranteed.