Brookeld Engineering Labs., Inc. Page 111 Manual No. M13-167-A0415
4) Therefore,ateachspeed,theacceptablewindowswithinwhichthemeasuredviscosityvalue
mustlieiscalculatedrelativetotheviscosityvalueofthestandard:
At6RPM:84.9cPto118.1cP
At12RPM:91.7cPto111.3cP
At30RPM:97.4cPto105.6cP
Ifyourmeasuredvaluesfalloutsideofthesewindows,contactBrookeldoryourauthorized
dealertodiscussyourresultsanddeterminewhetheryourinstrumentisoutofcalibration.