Brookeld Engineering Labs., Inc. Page 72 Manual No. M13-167-A0415
Administrator PowerUser User
YieldTestSetup Set
Spindle
ImmersionMark
Pre-ShearSpeed
Pre-ShearTime
WaitTime
RunSpeed
ZeroSpeed
TorqueReduction
YieldLimits
TemperatureSetpoint
GraphAxes
Instructions
SavePath
Table V-1: Factory Default User Settings
(continued from previous page)
V.4.3 Set Time and Date
TheTimeandDateisdisplayedontheStatusBaratthetopofeveryscreen.Theseparametersare
setwithintheSetTimeandDatemenu.TheformatforsettingTimeandDatewillbebasedonthe
RegionalSettings(seeSectionV.3).ChangefromSetTimetoSetDatebypressingthecommand
keyatthebottomofthescreen.WhenbothDateandTimearesetcorrectlyforyourlocation,press
OKtoacceptthechange.
V.4.4 Backup and Import
TheDV3TRheometerprovidesseveraloptionsrelatingtostorageofResults,TestsandUserSettings.
BackupandImportmenuoptionsrelatetocreatingbackuples,updatinginstrumentsoftware,
andupdatinglanguageles.
Backup:CreateabackuplefromtheDV3Tinternalmemoryincluding:Settings,UserProles,
Results,andTests.ThisbackupleissavedtoaUSBFlashDriveandcanbeuploaded(Restore)
toanyDV3TRheometer.
ChoosethelocationwithinaUSBFlashDrivewheretheBackupleistobesaved.Anewfolder
canbecreatedbyusingtheNewFolderbutton.PresstheBackupbuttontocreatetheBackuple.
Restore:UploadapreviouslysavedBackuple.TheinformationfromtheBackupleincludes:
Settings,UserProles,Results,andTests.ExistingUserproleswillnotbeoverwritten.User
prolesfromtheBackuplewillbeaddedtotheuserprolesthatalreadyexistontheDV3T.
LocatetheBackupleontheattachedUSBFlashDrive.PresstheRestorebuttontouploadthe
Backuple.
Automatic Backup:CreateBackuplesaccordingtoaschedule.ThefrequencyissettoDaily,
WeeklyorMonthlyincrements.ThePathmustbespeciedandrequirestheuseofaUSBFlash
Drive.ChecktheAllowAutomaticBackupcheckboxtoinitiatethisfeature.ThecurrentBackup
lewilloverwritethepreviousBackuplelocatedinthepathlocation.
TIP:YoucanpreservehistoricalbackuplesbyusingmultipleUSBFlashDrives.
Import:UploadSettingsfromaBackuple.Userproles,ResultsandTestsarenotincluded.
LocatetheBackupleontheUSBFlashDrive.PresstheImportbuttontouploadSettingsfrom
theBackuple.