Brookeld Engineering Labs., Inc. Page 65 Manual No. M13-167-A0415
Density g/cm
3
gram/cubiccentimeter
kg/m
3
kilogram/cubicmeter
1g/cm
3
=1000kg/m
3
TIP: Density and viscosity are both sensitive to
temperature. When using a density value in
CongureViscosityTest,besuretoenteradensity
valuethatwasdeterminedatthesametemperature
asthatoftheviscositymeasurement.
Regional Settings:TheRegionalSettingsmenudisplaysthecurrentselectionsforthedefault
language,numberformat,dateformatanddayseparator.Eachsettingcan
bechangedbypressingthebluedownarrowandselectingthedesiredvalue.
RegionalSettingsarepartofthedataleforsavedresults.
TheLanguagesettingestablishedthedefaultlanguagefortheDV3T.Individual
userscanchooseadifferentlanguageforusewiththeirspeciclogin(see
SectionV.2).
TheformatsforNumber,Time,DaySeparatorandDateofferseveraloptions.
NumberFormat 9,999,999.00 USAdecimalconvention
9.999.999,00 Europecommaconvention
9999999,00 SI/ISOspaceconvention
TimeFormat 13:59:59 24HourFormat
1:59:59PM 12HourFormat
DaySeparator / ForwardSlash
- Dash
. Period
DateFormat Year/Month/Day ISOStandard
Month/Day/Year USAStandard
Day/Month/Year InternationalStandard
Global Alarm: TheGlobalAlarmestablishesasinglemeasurementparameterwitharangeof
acceptance.Whenever the measured value exceedsthis rangeof acceptance,
anaudibleandvisualalarmwillbeissued.TheGlobalAlarmwillapplytoany
testthatisrun.TheQCLimitalsoestablishesameasurementparameterwitha
rangeofacceptance;however,theQCLimitisappliedonlytothetestinwhich
itisdened.
TheGlobalAlarmcanbesetforanyofthefourmeasurementparametersincluding;
ShearStress,Temperature,TorqueandViscosity.Onceaparameterisselected,
bothlowandhighlimitsmustbeentered.Theselimitsareindependentofthe
spindleandspeedselected.Whentheglobalalarmistriggered,theDV3Twill
provideanaudiblealarmandaWarningmessageonthescreen(seeFigureV-8).
Theerrormessagewillprovidesomedetailaboutthealarmcondition.Thetest