EasyManua.ls Logo

Bruker bravo - Laser status LED

Bruker bravo
72 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10
BRAVO Analyzer - User Manual
Safety 1
1.4.4 Laser status LED
The laser LED informs you about the current
laser status as follows:
Laser LED is off: Laser is disabled (i.e. it is
not ready for emitting radiation).
Blue solid: Laser is enabled (i.e. it is ready
for emitting radiation) and the laser safety
status of the instrument is OK.
Red solid: Laser emits radiation (for mea-
surement purposes). As soon as the mea-
surement is finished, the laser LED will
return to blue (solid).
Blue flashing: This indicates a laser safety
problem. In this case, you can not start a
measurement. Contact the Bruker service.
See section 1.10.
Laser status LED
Draft version

Table of Contents