10
BRAVO Analyzer - User Manual
Safety 1
1.4.4 Laser status LED
The laser LED informs you about the current
laser status as follows:
Laser LED is off: Laser is disabled (i.e. it is
not ready for emitting radiation).
Blue solid: Laser is enabled (i.e. it is ready
for emitting radiation) and the laser safety
status of the instrument is OK.
Red solid: Laser emits radiation (for mea-
surement purposes). As soon as the mea-
surement is finished, the laser LED will
return to blue (solid).
Blue flashing: This indicates a laser safety
problem. In this case, you can not start a
measurement. Contact the Bruker service.
See section 1.10.
Laser status LED