BRAVO Analyzer - User Manual
7 Troubleshooting
7.1 General information
This chapter deals with the most common problems that may occur as experience has
shown. It provides information about possible causes of the problem and presents solu-
tions for troubleshooting. If the solutions listed in this chapter do not solve the problem,
contact the Bruker service.
For the Bruker service contact data, see section 1.10.
7.2 Problem - possible cause - solution
7.2.1 BRAVO does not boot up
After being switched on, BRAVO does not boot up.
Possible causes Solutions
Empty battery Substitute the empty battery for a charged
battery.
See section 4.7.
Recharge the empty battery.
See section 4.8.
No battery is inserted in the BRAVO
analyzer.
Insert a charged battery.
See section 4.7.
A failed BRAVO software update
A defective partition on the SD card
Recover the system by pressing the ON/
OFF button and the laser enabling /
disabling button simultaneously.
A major BRAVO defect (e.g. damaged
touchscreen)
Contact the Bruker service.
For the service contact data, see
section 1.10. For detailed information
about the repair procedure, see
section 6.2.