EasyManua.ls Logo

Bruker bravo - Cleaning the touchscreen; Decontaminating the BRAVO analyzer

Bruker bravo
72 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
52
BRAVO Analyzer - User Manual Bruker Optik GmbH
Cleaning and Decontamination 5
5.2 Decontaminating the BRAVO analyzer
5.2.1 Approved decontaminating agents
The following detergents are approved for decontaminating the BRAVO analyzer:
Ethanol (C
2
H
6
O 70%)
Isopropanol (C
3
H
8
O 70%)
Sodium hydroxide (NaOH 5%)
Sodium hypochlorite (NaClO 6%)
Hydrogen peroxide (H
2
O
2
5%)
5.2.2 Decontamination procedure
1. Damp a cloth with an appropriate decontaminating agent. Use only an approved
decontaminating agent. (See the above list.)
2. Wipe off the BRAVO analyzer thoroughly.
3. Strictly observe the application time of the decontaminating agent in question.
(Note: The correct application is stated by manufacturer of the decontaminating
agent.)
4. Afterwards, remove the decontaminating agent by wiping off the BRAVO analyzer
thoroughly with a damp cloth (i.e. damped with water).
NOTICE
BRAVO analyzer damage because of improper decontamination method
Use only an approved decontaminating agent. See the above list. (Note: Only for
the above listed detergents, the material compatibility has been tested.)
Do not immerse the BRAVO analyzer into the liquid decontaminating agent.
Do not wash down the BRAVO analyzer.
Draft version

Table of Contents