Bruker Optik GmbH BRAVO Analyzer - User Manual
Troubleshooting 7
7.2.9 Problems regarding the docking station
You have put the BRAVO analyzer into docking station, but the blue lamp at the docking
station does not light up.
Possible causes Solutions
You have not put the BRAVO analyzer
properly into docking station.
Put the BRAVO analyzer again into the
docking station. Pay attention to the correct
sit of the contacts.
The corresponding contacts are dirty. Remove the dirt.
☞ For the location of the contact bars,
see section 3.1 and section 3.6.