Instruction Manual for Lenus Care Bed
Page 69
Inspection Report following an Inspection of Electromedical
Equipment according to DIN EN 62353 (VDE 0751-1): 2015-10 – Page 1 of 3
Customer / Medical facility / Practice:
Carried out: Repeat inspection Inspection prior to initial operation (reference value)
_____________________________________________ Inspection following repair/servicing
Equipment type: Hospital bed Care bed
Manufacturer: Burmeier & Co. KG
User-specific parts: Mattress base, headboard and footboard,
safety sides
Testing equipment used (type/inventory no.): 1.
Medical Devices Regulation classification: Class I
I. Visual inspection
What? How?
Visual inspection of the electrical components
Control unit housing, external plug-in
power adapter (optional equipment)
Correct position, damage, signs of
spilt liquids/contamination that may
affect the insulation
Housing and motor lifting tubes
Handset housing and keypad film
Motor cable, handset cable, mains
cable, connection cable
Plug and plug cover on control unit
Visual inspection of the mechanical components
Patient lifting pole, adapter sleeves;
grab handles
Bowden cable, CPR release,
backrest
Damage, splinters, dimensions acc.
to Sheet 3
Locking levers of safety sides, side
panels, headboard/footboard
Secure position of locking levers
Connecting elements (screws, bolts,
nuts, safety caps)
Secure fixing, missing parts
Wearing parts, such as joints
II. Electrical Measurement (use only measuring instruments according to DIN EN 62353 (VDE 0751-1))
Note: To minimise measuring errors, route the test leads as far away as possible from and not parallel to the power
cables and handset leads of the bed. Also observe the operating instructions for the measuring instruments used.
Insulation resistance (to be carried out only if there are doubts about the electrical insulation, such as:
If the customer’s RCD (residual current circuit breaker) has tripped several
times