Instruction Manual for Lenus Care Bed
Page 91
This device is compliant with the following EMC standards regarding interference emissions and
immunity:
Ambient limit values of the interference emissions
Professional healthcare facilities
Home healthcare environment
Conducted and radiated
interference emissions
Voltage fluctuations and
flicker
EMC basic
standard or
test method
Professional healthcare
facilities
Home healthcare environment
Electrostatic discharge (ESD)
+/- 2 kV, +/- 4 kV ¸ +/- 8 kV, +/- 15 kV air, +/- 25 kV air
High-frequency electromagnetic fields
3 V/m; 80 MHz up to 2.7 GHz;
80% AM at 1 kHz
10 V/m; (80 MHz up to 2.7 GHz; 80%
AM at 1 kHz)
High-frequency electromagnetic fields in the
immediate vicinity of wireless communication
devices
See separate table zz (at the end of this chapter)
Magnetic fields with rated power frequencies
Professional healthcare
facilities
Home healthcare environment
Electrical fast transients/ bursts
+/- 2 kV; 100 kHz repetition frequency
Electrical surges: conductor to conductor
Conducted interference induced by high-
frequency fields
3 V; 0.15 MHz to 80 MHz;
6 V in ISM frequency bands
between 0.15 MHz and
80 MHz
80% AM at 1 kHz
3 V; 0.15 MHz to 80 MHz;
6 V in ISM and amateur radio
frequency bands between 0.15 MHz
and 80 MHz
80% AM at 1 kHz
0% U
T
; ½ cycle;
at 0, 45, 90, 135, 180, 225, 270 and 315 degrees
0% U
T
; 1 cycle; and
70% U
T
; 25 cycles; single-phase at 0 degrees