Ports for signal input/signal output parts
Professional healthcare
facilities
Home healthcare
environment
Electrostatic discharge (ESD)
+/- 8 kV; contact
+/- 2 kV, +/- 4 kV, +/- 8 kV, +/- 15 kV; +/- 25 kV air
Electrical fast transients/ bursts
+/- 1 kV; 100 kHz repetition frequency
Conducted interference induced by high-
frequency fields
3 V; 0.15 MHz to 80 MHz;
6 V in ISM frequency bands
between 0.15 MHz and
80 MHz
80% AM at 1 kHz
3 V; 0.15 MHz to 80 MHz;
6 V in ISM and amateur radio
frequency bands between 0.15
MHz and 80 MHz
80% AM at 1 kHz
Table zz: Test specifications for the immunity of sheathings to high-frequency wireless
communication equipment
Immunity
test level
[v/m]
FM
+/- 5%
deviation,
1 kHz sine
wave
GSM 800/900
TETRA 800
iDEN820,
CDMA 850,
LTE band 5
GSM 1800
CDMA 1900,
GSM 1900,
DECT,
LTE band 1; 3;
4; 25; UMTS
Bluetooth,
WLAN 802.11
b/g/n,
RFID 2450,
LTE band 7