46
Troubleshooting
10. Remove the spill cup from the attenuator twisting counter-clockwise and then pull straight down.
11. Properly discard or liquid or contents that may be present in the spill cup.
12. Rinse the spill cup with water. If needed, wipe the spill cup with a clean, lint-free dry cloth.
13. Inspect the spill cup and lens for debris, scratching or cracks. If scratches or cracks are noticed, discard the
spill cup. Contact CEM Corporation or its authorized dealer if the spill cup has been damaged.
14. Inspect the inside of the cavity. If debris or liquid are located, remove the debris and wipe the cavity with a
clean, dry lint-free cloth. Contact CEM Corporation or its authorized dealer for additional support if debris or
liquid are noted in the cavity or on the IR lens.
15. Rinse the right outlet with water. The liquid will ow through the instrument and out of the trap. DO NOT place
any liquid into left inlet as this goes directly to the cavity blower.