GPIB/RS232 Interface (IEEE-488.2)
[:SOURce]:SAFety:FRAMe<f>:RESult:STEP<n>:MMETerage?
It queries the MEASURE METER readings of specified STEP for all channels of a
Frame when the Screen sets to On. The range of variable <f> at the end of FRAMe<f>
command is 0~9 that indicates the Frame address. When <f> is ignored the default is 1
it indicates to query the MEASURE METER readings of all channels for Slave 1.
Example: Input the command SAF:FRAME0:RES:STEP3:MMET?
The tester returns 7.000000E-05, 5.000000E-05, 4.000000E-05,
3.000000E-05, 2.000000E-05, 1.000000E-05, 2.000000E-04,
7.000000E-05, 5.000000E-04, 3.000000E-04.
Description: The returned result indicates the queried MEASURE METER
readings for all channels STEP3 of Address 0 host is 0.07mA,
0,05mA, 0.04mA, 0.03mA, 0.02mA, 0.01mA, 0.20mA, 0.07mA,
0.50mA, 0.30mA.
[:SOURce]:SAFety:FRAMe<f>:RESult:STEP<n>:OMETerage?
It queries the OUTPUT METER readings of specified STEP for all channels of a Frame
when the Screen sets to On. The range of variable <f> at the end of FRAMe<f>
command is 0~9 that indicates the Frame address. When <f> is ignored the default is 1
it indicates to query the OUTPUT METER readings of all channels for Slave 1.
Example: Input the command SAF:FRAM1:RES:STEP3:OMET?
The tester returns 5.100000E+01, 5.000000E+01, 5.000000E+01,
5.200000E+01, 5.000000E+01, 4.900000E+01, 5.100000E+01,
4.900000E+01, 5.000000E+01, 4.900000E+01.
Description: The returned result indicates the queried OUTPUT METER readings
for all channels STEP3 of Address 1 host is 0.051kV, 0.050kV,
0.050kV, 0.052kV, 0.050kV, 0.049kV, 0.051kV, 0.049kV, 0.050kV,
0.049kV.
[:SOURce]:SAFety:FRAMe<f>:RESult:STEP<n>:TIME[:ELAPsed]:DWELl?
It queries the DWELL time tested by specified STEP for all channels of a Frame when
the Screen sets to On. The range of variable <f> at the end of FRAMe<f> command is
0~9 that indicates the Frame address. When <f> is ignored the default is 1 it indicates to
query the DWELL time of all channels for Slave 1.
Example: Input the command SAF:FRAM0:RES:STEP1:TIME:DWEL?
The tester returns 3.000000E+00, 3.000000E+00, 3.000000E+00,
3.000000E+00, 3.000000E+00, 3.000000E+00, 3.000000E+00,
3.000000E+00, 3.000000E+00, 3.000000E+00.
Description: The returned result indicates the queried DWELL times tested by
STEP 1 for all channels of Address 0 host are 3sec, 3sec, 3sec, 3sec,
3sec, 3sec, 3sec, 3sec, 3sec, 3sec.
[:SOURce]:SAFety:FRAMe<f>:RESult:STEP<n>:TIME[:ELAPsed]:FALL?
It queries the time voltage fall elapsed of a specified STEP for all channels of a Frame
when the Screen sets to On. The range of variable <f> at the end of FRAMe<f>
command is 0~9 that indicates the Frame address. When <f> is ignored the default is 1
it indicates to query the time elapsed for voltage fall of all channels for Slave 1.
Example: Input the command SAF:FRAM0:RES:STEP4:TIME:FALL?
The tester returns 3.000000E+00, 3.000000E+00, 3.000000E+00,
3.000000E+00, 3.000000E+00, 3.000000E+00, 3.000000E+00,
3.000000E+00, 3.000000E+00, 3.000000E+00.