Multi-Channel Sync Hipot Tester 19020/19020-4/19021/19021-4/19022/19022-4
User’s Manual
Description: The returned result indicates the tested voltage fall times of STEP 4
for all channels of Address 0 host are 3sec, 3sec, 3sec, 3sec, 3sec,
3sec, 3sec, 3sec, 3sec, 3sec.
[:SOURce]:SAFety:FRAMe<f>:RESult:STEP<n>:TIME[:ELAPsed]:RAMP?
It queries the time voltage ramp elapsed of a specified STEP for all channels of a Frame
when the Screen sets to On. The range of variable <f> at the end of FRAMe<f>
command is 0~9 that indicates the Frame address. When <f> is ignored the default is 1
it indicates to query the time elapsed for voltage ramp of all channels for Slave 1.
Example: Input the command SAF:FRAM0:RES:STEP1:TIME:RAMP?
The tester returns 3.000000E+00, 3.000000E+00, 3.000000E+00,
3.000000E+00, 3.000000E+00, 3.000000E+00, 3.000000E+00,
3.000000E+00, 3.000000E+00, 3.000000E+00.
Description: The returned result indicates the time required for voltage to ramp up
to the set in STEP 1 for all channels of Address 0 host are 3sec, 3sec,
3sec, 3sec, 3sec, 3sec, 3sec, 3sec, 3sec, 3sec.
[:SOURce]:SAFety:FRAMe<f>:RESult:STEP<n>:TIME[:ELAPsed][:TEST]?
It queries the test time of a specified STEP for all channels of a Frame when the Screen
sets to On. The range of variable <f> at the end of FRAMe<f> command is 0~9 that
indicates the Frame address. When <f> is ignored the default is 1 it indicates to query
the test time of all channels for Slave 1.
Example: Input the command SAF:FRAM0:RES:STEP2:TIME?
The tester returns 3.000000E+00, 3.000000E+00, 3.000000E+00,
3.000000E+00, 3.000000E+00, 3.000000E+00, 3.000000E+00,
3.000000E+00, 3.000000E+00, 3.000000E+00.
Description: The returned result indicates the test time required for STEP 2 of all
channels for Address 0 host is 3sec, 3sec, 3sec, 3sec, 3sec, 3sec,
3sec, 3sec, 3sec, 3sec.
[:SOURce]:SAFety:STARt[:ONCE]
It starts the test.
Example: Input the command SAF:STAR
Description: It means to start the host test.
[:SOURce]:SAFety:STARt:CORRection:OPEN GET | OFF
When the parameter is set to GET, it gets the correction value, the host may output high
voltage. When the parameter is set to OFF, correction is disabled.
Example: Input the command SAF:STAR:CORR:OPEN GET
Description: It means to start the function of getting correction value.
[:SOURce]:SAFety:STARt:CORRection:OPEN?
It queries if correction has been done.
Example: Input the command SAF:STAR:CORR:OPEN?
The tester returns 0.
Description: The returned 0 means the correction hasn’t been done on the host.
[:SOURce]:SAFety:STARt:CORRection:SAMPle GET
It starts the GET Cs function in Open Short Check mode.