GPIB/RS232 Operation Description (IEEE-488.2)
5-35
Example: Input command “SAFE:STEP7:LC:CURR:OFFS:LDC?”.
Return message “+1.000000E-05”.
Description: Return message “+1.000000E-05” means the Offset value of the
human body impedance model in Step 7 is set to 10uA.
[:SOURce]:SAFEty:STEP<n>:LC[:DEVice] UL1950 | UL1563 | UL544NP |
UL544P | UL2601 | 1950-U1 | 2601-U1
This command sets which human body impedance model standards will be used
for the selected step.
Example: Input command “SAFE: STEP 7: LC: DEV UL544NP”
Description: This command sets the human body impedance model standard in
Step 7 to UL544NP.
[:SOURce]:SAFEty:STEP<n>:LC[:DEVice]?
This command queries which human body impedance model standards is being
used for the selected step.
Example: Input command “SAFE: STEP7: LC: DEV?”
Return message “UL544NP”
Description: Return message “UL544NP” means the human body impedance
model standard being used in STEP 7 is UL544NP.
[:SOURce]:SAFEty:STEP<n>:LC:DISPlay RMS|PEAK
This command sets the display mode for the LC leakage current for the selected
step.
Example: Input command “SAFE:STEP7:LC:DISPlay RMS”
Description: Sets the display mode for the LC leakage current in STEP 7 to RMS.
[:SOURce]:SAFEty:STEP<n>:LC:DISPlay?
This command queries the display mode for the LC leakage current for the
selected step.
Example: Input command “SAFE:STEP7:LC:DISPlay RMS”
Description: Return message “RMS” means the display mode for the LC leakage
current in STEP 7 is set to RMS.
[:SOURce]:SAFEty:STEP<n>:LC:LAC[:HIGH] <Range 0 ~ high limit> (0=OFF)
This command sets the high limit of the LAC leakage current in Amperes (A) for the
selected step.
Range: LAC HIGH LIMIT value ≤ LC HIGH LIMIT value (0=OFF)
DEVICE is UL544NP : 0.0000001∼0.006
DEVICE is UL544P : 0.0000001∼0.01
DEVICE is UL1563 : 0.0000001∼0.01
DEVICE is UL2601-1 : 0.0000001∼0.01
DEVICE is UL1950 : 0.0000001∼0.01
DEVICE is 1950-U1(RMS) : 0.0000001∼0.05
DEVICE is 1950-U1(PEAK) : 0.0000001∼0.07
DEVICE is 2601-U1 : 0.0000001∼0.01
Example: Input command “SAFE:STEP7:LC:LAC:HIGH 0.0001”
Description: This command sets the high limit of the LAC leakage current for STEP
7 to 0.1mA.
[:SOURce]:SAFEty:STEP<n>:LC:LAC?
This command queries the LAC leakage current value for the selected step.
Example: Input command “SAFE:STEP7:LC:LAC?