Electrical Safety Analyzer 19032 Users Manual
5-36
Description: Return message1.000000E-04means the high limit of the LAC
leakage current for STEP 7 is set to 0.1mA.
[:SOURce]:SAFEty:STEP<n>:LC:LDC[:HIGH] <Range 0 ~ high limit>0
(0=OFF)
This command sets the high limit of the LDC leakage current in Amperes (A) for the
selected step.
Range: 0 or 0.000001~HIGH LIMIT value. If the High Limit value is larger than 1mA
then the maximum value that can be set is 0.001 (0=OFF)
Example: Input command SAFE:STEP7:LC:LDC:HIGH 0.001
Description: This command sets the high limit of the LDC leakage current for step 7
to 1mA.
[:SOURce]:SAFEty:STEP<n>:LC:LDC?
This command queries the LDC leakage current value for the selected step.
Example: Input command SAFE:STEP7:LC:LDC?
Description: Return message1.000000E-03means the high limit of the LDC
leakage current for STEP 7 is set to 1mA.
[:SOURce]:SAFEty:STEP<n>:LC:LINE NORmal | REVerse | SFNormal |
SFReverse
This command selects which power circuit status simulation mode to use for the
selected step.
Example: Input command SAFE:STEP7:LC:LINE REV
Description: This command sets the power circuit status simulation mode for step 7
to Reverse.
[:SOURce]:SAFEty:STEP<n>:LC:LINE?
This command queries which power circuit status simulation mode is active for the
selected step.
Example: Input command SAFE:STEP7:LC:LINE?
Return message REVERSE
Description: Return message REVERSEmeans the power circuit status
simulation mode for STEP 7 is Reverse.
[:SOURce]:SAFEty:STEP<n>:LC:METEr L | P, P | G
This command sets the measurement point of leakage current for the selected step.
(The test mode in P-G isnt allowed.)
Example: Input command SAFE:STEP7:LC:METEr L,P
Description: This command sets the measurement point of leakage current of
STEP 7 to L, P.
[:SOURce]:SAFEty:STEP<n>:LC:METEr?
This command queries the measurement point of leakage current for the selected
step.
Example: Input command SAFE:STEP7:LC:METE?
Return message L, P
Description: Return message L, Pmeans the measurement point of leakage
current in STEP 7 is L-P.
[:SOURce]:SAFEty:STEP<n>:LC:GSWItch < boolean > | ON | OFF
This command sets the ground switch status for the selected step.
Example: Input command SAFE:STEP7:LC:GSWI ON
Description: This command sets the ground switch status for STEP 7 to ON.