Electrical Safety Analyzer 19032 User’s Manual
5-38
Example: Input command “SAFE:STEP7:LC:TIME 5”
Description: This command sets the test duration time for STEP 7 to 5.0sec.
[:SOURce]:SAFEty:STEP<n>:LC:TIME[:TEST]?
This command queries the test duration for the selected step.
Example: Input command “SAFE:STEP7:LC:TIME?”
Return message “5.000000E+00”
Description: Return message “5.000000E+00” means the test duration time for
STEP 7 is set to 5.0sec.
[:SOURce]:SAFEty:STEP<n>:LC:TIME:DWELl < numeric value >
This command sets the dwell time in seconds (s) for the selected step.
Range: 0 or 0.1~999.0. (0=OFF).
Example: Input command “SAFE:STEP7:LC:TIME:DWEL 2.0”
Description: This command sets the dwell time for STEP 7 to 2.0sec.
[:SOURce]:SAFEty:STEP<n>:LC:TIME:DWELl?
This command queries the dwell time for the selected step.
Example: Input command “SAFE:STEP 7:LC:TIME:DWEL?”
Return message “2.000000E+00”
Description: Return message “2.000000E+00” means the dwell time setting for
STEP 7 is set to 2.0sec.
[:SOURce]:SAFEty:STEP<n>:LC:POWer:MODE VOLTage | CURRent | VA |
SIMUlation | SOURce
This command sets the power measurement mode for the selected step.
Example: Input command “SAFE:STEP7:LC:POW:MODE VOLTage”
Description: This command sets the power measurement mode for STEP 7 to
VOLTAGE.
[:SOURce]:SAFEty:STEP<n>:LC:POWer:MODE?
This command queries the power measurement mode for the selected step.
Example: Input command “SAFE:STEP 7:LC:POW: MODE?”
Return message “VOLTAG E”
Description: Return message “VOLTAGE” means the power measurement mode
for STEP 7 is set to VOLTAGE.
[:SOURce]:SAFEty:STEP<n>:LC:POWer:VOLTage[:LIMit][:HIGH] < numeric
value >
This command sets the power voltage measurement high limit in volts (V) for the
selected step.
Range: 0 or 0.1~300.0. (0=OFF).
Example: Input command “SAFE:STEP7:LC:POW:VOLT 220”
Description: This command sets the power voltage measurement high limit value
for STEP 7 to 220V.
[:SOURce]:SAFEty:STEP<n>:LC:POWer:VOLTage[:LIMit][:HIGH]?
This command queries the power voltage measurement high limit value for the
selected step.
Example: Input command “SAFE:STEP7:LC:POW:VOLT?”
Return message “2.200000E+02”
Description: Return message “2.200000E+02” means the power voltage
measurement high limit value for STEP 7 is set to 220V.