GPIB/RS232 Operation Description (IEEE-488.2)
5-39
[:SOURce]:SAFEty:STEP<n>:LC:POWer:VOLTage[:LIMit]:LOW < numeric
value >
This command sets the power voltage measurement low limit value in volts (V).
Range: 0 or 0.1~300.0. (0=OFF) (Low limit value of voltage measurement high
limit value)
Example: Input command SAFE:STEP7:LC:POW:VOLT:LOW 110
Description: This command sets the power voltage measurement low limit value for
STEP 7 to 110V.
[:SOURce]:SAFEty:STEP<n>:LC:POWer:VOLTage[:LIMit]:LOW?
This command queries the power voltage measurement low limit value for the
selected step.
Example: Input command SAFE:STEP7:LC:POW:VOLT:LOW?
Return message 1.100000E+02
Description: Return message 1.100000E+02 means the power voltage
measurement low limit value for STEP 7 is set to 110V.
[:SOURce]:SAFEty:STEP<n>:LC:POWer:CURRent[:LIMit][:HIGH] < numeric
value >
This command sets the power current measurement high limit value in amperes (A)
for the selected step.
Range: 0 or 0.001~20. (0=OFF).
Example: Input command SAFE:STEP7:LC:POW:CORR 5
Description: This command sets the power current measurement high limit value
for STEP 7 to 5A.
[:SOURce]:SAFEty:STEP<n>:LC:POWer:CURRent[:LIMit][:HIGH]?
This command queries the power current measurement high limit value for the
selected step.
Example: Input command SAFE:STEP7:LC:POW:CURR?
Return message 5.000000E+00
Description: Return message 5.000000E+00means the power current
measurement high limit value for STEP 7 is set to 5A.
[:SOURce]:SAFEty:STEP<n>:LC:POWer:CURRent[:LIMit]:LOW < numeric
value >
This command sets the power current measurement low limit value in amperes (A)
for the selected step.
Range: 0 or 0.001~20. (0=OFF). (Low limit value of current measurement high
limit value).
Example: Input command SAFE:STEP7:LC:POW:CORR:LOW 0.5
Description: This command sets the power current measurement low limit value for
STEP 7 to 0.5A.
[:SOURce]:SAFEty:STEP<n>:LC:POWer:CURRent[:LIMit]:LOW?
This command queries the power current measurement low limit value for the
selected step.
Example: Input command SAFE:STEP7:LC:POW:CURR:LOW?
Return message 5.000000E-01
Description: Return message 5.000000E-01means the power current
measurement low limit value for STEP 7 is set to 0.5A.