Electrical Safety Analyzer 19032 Users Manual
5-40
[:SOURce]:SAFEty:STEP<n>:LC:POWer:VA[:LIMit][:HIGH] < numeric value >
This command sets the power measurement high limit value In volt-amperes (VA)
for the selected step.
Range: 0 or 0.1~4400. (0=OFF).
Example: Input command SAFE:STEP7:LC:POW:VA 110
Description: This command sets the power measurement high limit value for STEP
7 to 110VA.
[:SOURce]:SAFEty:STEP<n>:LC:POWer:VA[:LIMit][:HIGH]?
This command queries the power measurement high limit value for the selected
step.
Example: Input command SAFE:STEP7:LC:POW:VA?
Return message 1.100000E+02
Description: Return message 1.100000E+02 means the power measurement
high limit value for STEP 7 is set to 110VA.
[:SOURce]:SAFEty:STEP<n>:LC:POWer:VA[:LIMit]:LOW < numeric value >
This command sets the power measurement low limit value step in volt-amperes
(VA) for the selected.
Range: 0 or 0.1~4400. (0=OFF). (Low limit value of power measurement high
limit value)
Example: Input command SAFE:STEP7:LC:POW:VA:LOW 90
Description: This command sets the power measurement low limit value for STEP
7 to 90VA.
[:SOURce]:SAFEty:STEP<n>:LC:POWer:VA[:LIMit]:LOW?
This command queries the power measurement low limit value for the selected
step.
Example: Input command SAFE:STEP7:LC:POW:LOW:VA?
Return message 9.000000E+01
Description: Return message 9.000000E+01 means the power measurement
low limit value for STEP 7 is set to 90VA.
[:SOURce]:SAFEty:STEP<n>:LC:POWer:SIMUlation:TVOLtage < numeric
value >
This command sets the target value of the power voltage simulation in volts (V) for
the selected step.
Range: 80~300
Example: Input command SAFE:STEP 7:LC:POW:SIMU:TVOL 127
Description: This command sets the target value of the power voltage simulation
for STEP 7 to 127V.
[:SOURce]:SAFEty:STEP<n>:LC:POWer:SIMUlation:TVOLtage?
This command queries the target value of the power voltage simulation for the
selected step.
Example: Input command SAFE:STEP7:LC:POW:SIMU:TVOL?
Return message 1.270000E+02
Description: Return message 1.270000E+02means the target value of the
power voltage simulation for STEP 7 is set to 127V.
[:SOURce]:SAFEty:STEP<n>:LC:POWer:SOURce:TVOLtage < numeric value >
This command sets the source power voltage value in volts (V) for the selected
step.
Range: 80~300