GPIB/RS232 Operation Description (IEEE-488.2)
5-41
Example: Input command “SAFE:STEP7:LC:POW:SOUR:TVOL 90”
Description: This command sets the source power voltage value for STEP 7 to 90V.
[:SOURce]:SAFEty:STEP<n>:LC:POWer:SOURce:TVOLtage?
This command queries the source power voltage value for the selected step.
Example: Input command “SAFE:STEP7:LC:POW:SOUR:TVOL?”
Return message “9.000000E+01”
Description: Return message “9.000000E+01” means the source power voltage
value for STEP 7 is set to 90V.
[:SOURce]:SAFEty:STEP<n>:LC:POWer:SOURce: TFRequency< numeric value >
This command sets the source power frequency value in hertz (Hz) for the selected
step.
Range: 50~60
Example: Input command “SAFE:STEP7:LC:POW:SOUR:TFR 60”
Description: This command sets the source power frequency value for STEP 7 to
60Hz.
[:SOURce]:SAFEty:STEP<n>:LC:POWer:SOURce: TFRequency?
This command queries the source power frequency value for the selected step.
Example: Input command “SAFE:STEP7:LC:POW:SOUR:TFR?”
Return message “6.000000E+01”
Description: Return message “6.000000E+01” means the source power frequency
value for STEP 7 is set to 60Hz.
[:SOURce]:SAFEty:STEP<n>:LC:UPM < boolean > | ON | OFF
This command sets the UPM status of the selected step.
Example: Input command “SAFE:STEP7:LC:UPM ON”
Description: The command sets the UPM status for STEP 7 to ON.
[:SOURce]:SAFEty:STEP<n>:LC:UPM?
This command queries the UPM status of the selected step.
Example: Input command “SAFE:STEP7:LC:UPM?”
Return message “1”
Description: Return message “1” means the UPM status for STEP 7 is set to ON.
[:SOURce]:SAFEty:STEP<n>:LC:CHANnel[:LOW] < channel list >
This command sets selected step which setting of return terminal.
Example: Input command “SAFE:STEP7:LC:CHAN (@1(3))”
Description: This command sets output channel status of scanning test of the main
unit STEP 7 to BOX 1 channel 3 LOW output.
[:SOURce]:SAFEty:STEP<n>:LC:CHANnel[:LOW]?
This command queries selected step which setting of return terminal.
Example: Input command “SAFE:STEP7:LC:CHAN?”
Return message “(@1(3))”
Description: Return message “(@1(3))” means output channel status of scanning
test STEP 7 is BOX 1 channel 3 LOW output.
[:SOURce]:SAFEty:PRESet:TIME:PASS < numeric value >
This command sets the length of time the buzzer sounds in seconds (s) when a test
passes.
Range: 0.2~99.9.
Example: Input command “SAFE:PRES:TIME:PASS 3”