Electrical Safety Analyzer 19032 Users Manual
5-42
Description: This command sets the length of time the buzzer sounds to 3 seconds
when the test passes.
[:SOURce]:SAFEty:PRESet:TIME:PASS?
This command queries the length of time the buzzer sounds when a test passes.
Example: Input command SAFE:PRES:TIME:PASS?
Return message 3.000000E+00
Description: Return message 3.000000E+00 means the buzzer sound time is 3
seconds when the test passes.
[:SOURce]:SAFEty:PRESet:TIME:STEP < numeric value > | KEY
This command sets the interval time between steps, or from the next start
command to execute the next step, in seconds (s).
Range: KEY or 0.199.9.
Example: Input command SAFE:PRES:TIME:STEP 0.5
Description: This command sets the interval time between steps to 0.5 seconds.
[:SOURce]:SAFEty:PRESet:TIME:STEP?
This command queries the interval time between steps.
Example: Input command SAFE:PRES:TIME:PASS?
Return message 5.000000E-01
Description: Return message 5.000000E-01means the interval time between
steps is set to 0.5 second.
[:SOURce]:SAFEty:PRESet:TIME:ASTart < numeric value >
This command sets the Start Wait delay time in seconds (s).
Range: 0 or 0.1~99.9. (0=OFF).
Example: Input command SAFE:PRES:TIME:AST 1
Description: This command sets the Start Wait delay time to 1 second.
[:SOURce]:SAFEty:PRESet:TIME:ASTart?
This command queries the Start Wait delay time.
Example: Input command SAFE:PRES:TIME:AST?
Return message 1.000000E+00
Description: Return message 1.000000E+00 means the Start Wait delay time is
set to 0.5 seconds.
[:SOURce]:SAFEty:PRESet:GB:FREQuency < numeric value >
This command sets the output current frequency during ground bond testing in
Hertz (Hz).
Range: 50/60
Example: Input command SAFE:PRES:GB:FREQ 50
Description: This command sets the output current frequency to 50Hz during
ground bond testing.
[:SOURce]:SAFEty:PRESet:GB:FREQuency?
This command queries the output current frequency during ground bond testing.
Example: Input command SAFE:PRES:GB:FREQ?
Return message 5.000000E+01
Description: Return message 5.000000E+01means the output current
frequency is set to 50Hz during ground bond testing.
[:SOURce]:SAFEty:PRESet:GB:VOLTage < numeric value >
This command sets the open circuit voltage during ground bond testing, in Volts