GPIB/RS232 Operation Description (IEEE-488.2)
5-43
(V).
Range: 6~9
Example: Input command “SAFE:PRES:GB:VOLT 9”
Description: This command sets the open circuit voltage to 9V during ground bond
testing.
[:SOURce]:SAFEty:PRESet:GB:VOLTage?
This command queries the open circuit voltage during ground bond testing.
Example: Input command “SAFE:PRES:GB:VOLT?”
Return message “9.000000E+00”
Description: Return message “9.000000E+00” means the open circuit voltage is
set to 9V during ground bond testing.
[:SOURce]:SAFEty:PRESet:AC:FREQuency < numeric value >
This command sets the output voltage frequency for the AC withstand
voltage test in hertz (Hz).
Range: 50~600
Example: Input command “SAFE:PRES:AC:FREQ 60”
Description: This command sets the output voltage frequency to 60Hz for the
AC withstand voltage test.
[:SOURce]:SAFEty:PRESet:AC:FREQuency?
This command queries the output voltage frequency for the AC withstand
Voltage test.
Example: Input command “SAFE:PRES:AC:FREQ?”
Return message “6.000000E+01”
Description: Return message “6.000000E+01” means the output voltage
frequency is set to 60Hz for the AC withstand voltage test.
[:SOURce]:SAFEty:PRESet:WRANge[:AUTO] < boolean > | ON | OFF
This command sets the withstand voltage auto range function to O N or O FF.
Example: Input command “SAFE:PRES:WRAN ON”
Description: This command sets the withstand voltage auto range function to ON.
[:SOURce]:SAFEty:PRESet:WRANge[:AUTO]?
This command queries if the withstand voltage auto range function is ON or OFF.
Example: Input command “SAFE:PRES:WARN?”
Return message “1”
Description: Return message “1” means the withstand voltage auto range function
is ON.
[:SOURce]:SAFEty:PRESet:AGC[:SOFTware] < boolean > | ON | OFF
This command sets the software AGC to ON or OFF.
Example: Input command “SAFE:PRES:AGC ON”
Description: This command sets the software AGC to ON.
[:SOURce]:SAFEty:PRESet:AGC[:SOFTware]?
This command queries if the software AGC is ON or OFF.
Example: Input command “SAFE:PRES:AGC?”
Return message “1”
Description: Return message “1” means the software AGC is ON.
[:SOURce]:SAFEty:PRESet:NUMber:PART < string data >
This command sets the part number of the product.