Electrical Safety Analyzer 19032 User’s Manual
4-2
Display Area
Title List: This list displays the current setting of the analyzer or testing mode.
Function Key Message Display area:
Different function descriptions are displayed for each of the different menus.
The right side of the display shows the corresponding Function Keys. If the
description is blank or gray scale font, the corresponding function is invalid.
Message Bar: This list displays the setting mode, the range of the setting value, and the
testing time.
Message Indication Diagram:
Remote: When this area is highlighted, the analyzer is under remote control (the
analyzer is being controlled by a PC through either the RS-232 or GPIB
interfaces connected to the PC). At the same time, all of the keys are
disabled except for the [STOP] and [LOCAL] keys.
Lock: When this area is highlighted, the analyzer is in parameter-protected mode.
All other keys are disabled except for “MEMORY”, ”TEST”, “KEY LOCK”, and
“ERROR LOG” modes.
Offset: When this area is highlighted, the analyzer has zeroed the leakage current of
test cables and test leads.
Error: When this area is highlighted, an error message has been received.