Electrical Safety Analyzer 19032 User’s Manual
4-4
(8) Cal-Enable: Calibration switch. Used by qualified service personnel for unit
calibration. Should NOT be used by any other operator.
(9) Update: Update firmware switch. Used by qualified service personnel for unit
service. Should NOT be used by any other operator.
(10) Data Entry Keys/Program Key
[0][.]
[9]: Numeric key/character key, for inputting test parameter data (numeral
or alphabet). Under “MAIN MENU” display mode, [1], [2], [3], [4], [5]
keys can enter each display mode.
[ENTER]: Enter key. After entering a test parameter, press this key to confirm
entry.
[CLR]: Clear key. If an error is made when entering a test parameter, press
this key to cancel the input.
(11) Function Keys: Function Key. Selects sub-menus and enters test parameters. If the
description is blank, the corresponding function is invalid.
Terminal Area
(12) OUTPUT: High Voltage Terminal (AC/DC Hipot & IR).
This terminal has a high electric potential output. Do not touch when
the DANGER LED is illuminated.
(13) DRIVE (+): High Current Terminal (Driver) for Ground Bond Test.
(14) SENSE (+): High Voltage Terminal (Sense) for Ground Bond Test
(15) SENSE (-): Low Voltage Terminal (Sense) for Ground Bond Test.
(16) DRIVE (-): Ground Reference for ALL Tests.
This terminal is the high voltage test and reference terminal ground
(low electric potential terminal).