GPIB/RS232 Operation Description (IEEE-488.2)
5-13
The time left in the ramp cycle
The time left in the dwell cycle.
The elapsed time of test.
Return the leave time when Test Time is limited.
Return 9.9000001E+37 when Test Time is CONT.
The time left in the test cycle.
Return the leave time when Test Time is limited.
Return 9.9000001E+37 when Test Time is CONT.
The current elapsed Fall Time.
The current remaining Fall Time.
Example: Input command “SAFE: FETH?”STEP, MODE, OMET
Return message “1, AC, +5.000000E+02”
Description: Return message “1, AC, +5.000000E+02” means the current
STEP, MODE, and output values are set to STEP1, AC MODE and
0.500kV.
[:SOURce]:SAFEty:STARt[:ONCE]
This command is for starting the test.
Example: Input command “SAFE:STAR”
Description: This command starts the test.
[:SOURce]:SAFEty:STARt:OFFSet GET | OFF
This command gets the offset value when the parameter is GET and turns off the
offset function when the parameter is OFF.
Example: Input command “SAFE:STAR OFFS GET”
Description: This command gets the offset value.
Example: Input command “SAFE:STAR OFFS OFF”
Description: This command turns the offset function off.
[:SOURce]:SAFEty:STARt:OFFSet?
This command queries if the offset function is ON or OFF
Example: Input command “SAFE:STAR OFFS?”
Return message “0”
Description: Return message “0” means the analyzer is not using the offset
function
[:SOURce]:SAFEty: STARt: CSTandard GET
This command triggers the unit to get the Cs value
Example: Input command “SAFE: STAR: CST GET”
Description: This command starts the GET Cs function for the open/short circuit
detection mode. A known good DUT must be connected when this
command is sent.
[:SOURce]:SAFEty:STOP
This command is for stopping the test.
Example: Input command “SAFE:STOP”
Description: This command stops the analyzer test.