EasyManua.ls Logo

Chroma 19032 - Page 80

Default Icon
138 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Electrical Safety Analyzer 19032 Users Manual
5-14
:SOURce]:SAFEty:BCONtinue
This command is able to interrupt the step testing when the TEST TIME of the
testing step is set to CONTINUE.
Example: Input command SAFE:BCON
Description: It interrupts the step testing.
[:SOURce]:SAFEty:STATus?
This command queries the execution status of the current device under test. The
returned character data is either RUNNING or STOPPED.
Example: Input command SAFE:STAT?
Return message RUNNING
Description: Return message RUNNINGmeans the analyzer is running the test.
[:SOURce]:SAFEty:SNUMber?
This command queries how many steps have been set in the memory.
Example : Input command SAFE:SNUM?
Return message +2
Description: Return message +2 means 2 steps in the main memory have been
set.
[:SOURce]:SAFEty:RESult:ALL:OMETerage?
This command queries all the voltage (OUTPUT METER) readings for the step.
Example : Input commandSAFE:RES:ALL:OMET?”
Return message 5.100000E+01
Description: Return message 5.100000E+01 means the query OUTPUT METER
returned a value of 0.051kV.
[:SOURce]:SAFEty:RESult:ALL:MMETerage[:NORMal]?
This command queries all the MEASURE METER readings for the step.
Example : Input command SAFE:RES:ALL:MMET?”
Return message 7.000000E-05
Description: Return message 7.000000E-05 means the query MEASURE
METER returned a value of 0.07mA.
[:SOURce]:SAFEty:RESult:ALL:MMETerage:LAC?
This command queries all the MEASURE METER LAC readings for the step.
Example : Input commandSAFE:RES:ALL:MMET:LAC?
Return message1.000000E-04
Description: Return message 1.000000E-04means the query MEASURE
METER LAC returned a value of 0.10mA.
[:SOURce]:SAFEty:RESult:ALL:MMETerage:LDC?
This command queries all the MEASURE METER LDC readings for the step.
Example : Input command SAFE:RES:ALL:MMET:LDC?
Return message 1.000000E-04
Description: Return message 1.000000E-04means the query MEASURE
METER LDC returned a value of 0.10mA.
[:SOURce]:SAFEty:RESult:ALL:MODE?
This command queries the MODE of all steps. The returned format is:
AC|DC|GB|IR|LC|PA|OSC.
Example: Input command SAFE:RES:ALL:MODE?
Return message DC
Description: Return message DCmeans the unit is set to test mode DC.

Table of Contents

Other manuals for Chroma 19032

Related product manuals