Electrical Safety Analyzer 19032 User’s Manual
5-14
:SOURce]:SAFEty:BCONtinue
This command is able to interrupt the step testing when the TEST TIME of the
testing step is set to CONTINUE.
Example: Input command “SAFE:BCON”
Description: It interrupts the step testing.
[:SOURce]:SAFEty:STATus?
This command queries the execution status of the current device under test. The
returned character data is either RUNNING or STOPPED.
Example: Input command “SAFE:STAT?”
Return message “RUNNING”
Description: Return message “RUNNING” means the analyzer is running the test.
[:SOURce]:SAFEty:SNUMber?
This command queries how many steps have been set in the memory.
Example : Input command “SAFE:SNUM?”
Return message “+2”
Description: Return message “+2” means 2 steps in the main memory have been
set.
[:SOURce]:SAFEty:RESult:ALL:OMETerage?
This command queries all the voltage (OUTPUT METER) readings for the step.
Example : Input command “SAFE:RES:ALL:OMET?”
Return message “5.100000E+01”
Description: Return message “5.100000E+01” means the query OUTPUT METER
returned a value of 0.051kV.
[:SOURce]:SAFEty:RESult:ALL:MMETerage[:NORMal]?
This command queries all the MEASURE METER readings for the step.
Example : Input command “SAFE:RES:ALL:MMET?”
Return message “7.000000E-05”
Description: Return message “7.000000E-05” means the query MEASURE
METER returned a value of 0.07mA.
[:SOURce]:SAFEty:RESult:ALL:MMETerage:LAC?
This command queries all the MEASURE METER LAC readings for the step.
Example : Input command “SAFE:RES:ALL:MMET:LAC?”
Return message “1.000000E-04”
Description: Return message “1.000000E-04” means the query MEASURE
METER LAC returned a value of 0.10mA.
[:SOURce]:SAFEty:RESult:ALL:MMETerage:LDC?
This command queries all the MEASURE METER LDC readings for the step.
Example : Input command “SAFE:RES:ALL:MMET:LDC?”
Return message “1.000000E-04”
Description: Return message “1.000000E-04” means the query MEASURE
METER LDC returned a value of 0.10mA.
[:SOURce]:SAFEty:RESult:ALL:MODE?
This command queries the MODE of all steps. The returned format is:
AC|DC|GB|IR|LC|PA|OSC.
Example: Input command “SAFE:RES:ALL:MODE?”
Return message “DC”
Description: Return message “DC” means the unit is set to test mode DC.