EasyManua.ls Logo

Chroma 19032 - Page 82

Default Icon
138 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Electrical Safety Analyzer 19032 Users Manual
5-16
POWER LOW
FAIL
59
89
SHORT FAIL
61
97
OPEN FAIL
62
98
Example : Input command SAFE:RES:ALL?
Return message 116
Description : Return message 116 means judgment result is pass.
[:SOURce]:SAFEty:RESult:COMPleted?
This command queries if all the test results are completed. Returns 1 or 0.
Example: Input command SAFE:RES:COMP?
Return message 1”
Description: Return message 1” means all of the test results are completed.
[:SOURce]:SAFEty:RESult[:LAST][:JUDGment]?
This command queries the judgment result code of the last step.
Example: Input command SAFE:RES:LAST?
Return message 116
Description: This command means the judgment result of the last step is pass.
[:SOURce]:SAFEty:RESult:AREPort < boolean > | ON | OFF
This command sets the automatic reporting of the test results (RS-232 Interface
only).
Example: Input command SAFE:RES:AREP ON
Description: The analyzer auto reports the test result after the test is completed.
[:SOURce]:SAFEty:RESult:AREPort?
This command queries if the Automatic Reporting of test results is ON or OFF.
Returns 1 or 0 (RS-232 interface only).
Example: Input command SAFE:RES:AREP?
Return message 1”
Description: Return message 1” means auto report test result after the analyzer
test is completed is “ON”.
[:SOURce]:SAFEty:RESult:AREPort:ITEM [ < item > ] [ , < item > ]
Sets test data for auto report; < item > is for character data (meaning is listed
below):
Character Data
Returned Data
MODE
Measurement MODE.
OMETer ag e
Output value
MMETerage
Measurement value
LACMETerage
LAC measurement value
LDCMETerage
LDC measurement value
RELApsed
The elapsed time for Ramp.
DELApsed
The elapsed time for Dwell.
TELApsed
The elapsed time for Test. When Test Time is set to CONT.
and Test Time is higher than 999sec, it responds
9.9000001E+37.
FELApsed
The elapsed time for Fall Time.
STATe
The test result code