Electrical Safety Analyzer 19032 User’s Manual
5-20
[:SOURce]:SAFEty:STEP<n>:GB:CHANnel[:HIGH]?
This command queries the high scan channels for the selected step.
Example : Input command “SAFE:STEP1:AC:CHAN?”
Return message “(@2(1,2))”
Description : Return message “(@2(1,2))” means the high scan channel in
STEP 1 is set to BOX 2 channel 1 and 2 HIGH output.
[:SOURce]:SAFEty:STEP<n>:AC:CURRent>:OFFSet <numeric value>
This command sets the Offset value of the AC withstand test of the selected step.
Parameter <numeric value>: The Offset value of the AC withstand test in
Amperes (A).
Range: The Offset range is 0.000000~0.002999 when High Limit range is from
0.001 to 2.999mA.
The Offset range is 0.00000~0.04000 when High Limit range is from 3 to
40mA.
Example: Input command “SAFE:STEP2:AC:CURR:OFFS 0.0001”.
Description: Sets the Offset value of the AC withstand test in Step 2 of the
analyzer to 0.1mA.
[:SOURce]:SAFEty:STEP<n>:AC:CURR:OFFSet?
This command queries the Offset value of the AC withstand test of the selected
step.
Example: Input command “SAFE:STEP2:AC:CURR:OFFS?”.
Return message “+1.000000E-04”
Description: Return message “+1.000000E-04” means the Offset value of the AC
withstand test in Step 1 is 0.1mA.
[:SOURce]:SAFEty:STEP<n>:AC:FREQuency <numeric value>
This command sets the test frequency of the AC withstand test for the selected
step.
Parameter <numeric value>: The test frequency of the AC withstand test in Hertz
(Hz).
Range: 0 or 50~600Hz (0=DEFAULT). When the setting is 0, the output
frequency is set by the AC Freq. setting under PRESET.
Example: Input command “SAFE:STEP2:AC:FREQ 60”
Description: Sets the frequency of the AC withstand test in Step 2 of the main
unit to 60Hz.
[:SOURce]:SAFEty:STEP<n>:AC:FREQuency?
This command queries the test frequency of the AC withstand test for the selected
step.
Example: Input command “SAFE:STEP2:AC:FREQ?”
Return message “+6.000000E+01”
Description: Return message “+6.000000E+01” means the test frequency of the
AC withstand test in Step 2 is 60Hz.
[:SOURce]:SAFEty:STEP<n>:AC[:LEVel]< numeric value >
This command sets the AC withstand voltage level in volts (V) for the selected step.
Range: 50~5000
Example: Input command “SAFE:STEP2:AC 3000”
Description: Sets the AC withstand voltage level in STEP 2 to 3000V.