GPIB/RS232 Operation Description (IEEE-488.2)
5-21
[:SOURce]:SAFEty:STEP<n>:AC[:LEVel]?
This command queries the AC withstand voltage level for the selected step.
Example: Input command SAFE:STEP2:AC?
Return message 3.000000E+03
Description: Return message 3.000000E+03 means the voltage in Step 2 of the
AC withstand voltage test is 3000V.
[:SOURce]:SAFEty:STEP<n>:AC:LIMit[:HIGH] < numeric value >
This command sets the AC withstand voltage leakage current high limit in
amperes (A) for the selected step.
Range: 0.000001~0.04
Example: Input command SAFE:STEP2:AC:LIM 0.01
Description: This command sets the AC withstand voltage leakage current high limit
in STEP 2 to 10mA.
[:SOURce]:SAFEty:STEP<n>:AC:LIMit[:HIGH]?
This command queries the AC withstand voltage leakage current high limit value for
the selected step.
Example : Input command SAFE:STEP2:AC:LIM?
Return message 1.000000E-02
Description: Return message 1.000000E-02means the AC withstand voltage
leakage current high limit value in STEP 2 is 10mA.
[:SOURce]:SAFEty:STEP<n>:AC:LIMit:LOW < numeric value >
This command sets the AC withstand voltage leakage current low limit value for the
selected step.
Example : Input command SAFE:STEP2:AC:LIM:LOW 0.00001
Description: This command sets the AC withstand voltage leakage current low limit
value in STEP 2 to 0.01mA.
[:SOURce]:SAFEty:STEP<n>:AC:LIMit:LOW?
This command queries the AC withstand voltage leakage current low limit value for
the selected step.
Range : 0.000001~0.04 (low limit value of leakage current setting high limit
value)
Example : Input command SAFE:STEP2:AC:LIM:LOW?
Return message 1.000000E-05
Description: Return message 1.000000E-05means the AC withstand voltage
leakage current low limit value in STEP 2 is 0.01mA.
[:SOURce]:SAFEty:STEP<n>:AC:LIMit:ARC[:LEVel] < numeric value >
This command sets the arc current limit in amperes (A) for the selected step.
Range : 0 or 0.0010~0.0300 (0=OFF).
Example : Input command SAFE:STEP2:AC:LIM:ARC 0.004
Description: This command sets the arc current limit in STEP 2 to 4mA.
[:SOURce]:SAFEty:STEP<n>:AC:LIMit:ARC[:LEVel]?
This command queries the arc current limit for the selected step.
Example : Input command SAFE:STEP2:AC:LIM:ARC?
Return message 4.000000E-03
Description: Return message 4.000000E-03 means the arc current limit in STEP
2 is 4.0mA.