EasyManua.ls Logo

Chroma 19032 - Page 88

Default Icon
138 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Electrical Safety Analyzer 19032 Users Manual
5-22
[:SOURce]:SAFEty:STEP<n>:AC:LIMit:ARC:FILTer < numeric value >
This command sets the arc filter in Hertz (Hz) for the selected step.
Range : ARC bandwidth 23 kHz = 2.300000E+04
ARC bandwidth 50 kHz = 5.000000E+04
ARC bandwidth 100 kHz = 1.000000E+05
ARC bandwidth 230 kHz = 2.300000E+05
Example : Input commandSAFE:STEP2:AC:LIM:ARC:FILT 230000
Description : This command sets the arc filter in STEP 2 to 230000Hz (230kHz).
[:SOURce]:SAFEty:STEP<n>:AC:LIMit:ARC:FILTer?
This command queries the arc filter for the selected step.
Example: Input command SAFE:STEP2:AC:LIM:ARC:FILT?
Return message 2.300000E+05
Description: Return message 2.300000E+05 means the arc filter in STEP 2 is
230000Hz (230kHz).
[:SOURce]:SAFEty:STEP<n>:AC:TIME:RAMP < numeric value >
This command sets the ramp time in seconds (s) for the selected step.
Range : 0 or 0.1~999.0 (0=OFF).
Example : Input command SAFE:STEP2:AC:TIME:RAMP 5
Description: This command sets the ramp time in STEP 2 to 5.0sec.
[:SOURce]:SAFEty:STEP<n>:AC:TIME:RAMP?
This command queries the ramp time for the selected step.
Example: Input command SAFE:STEP2:AC:TIME:RAMP?
Return message 5.000000E+00
Description: Return message 5.000000E+00 means the ramp time in STEP 2 is
5.0sec.
[:SOURce]:SAFEty:STEP<n>:AC:TIME[:TEST] < numeric value >
This command sets the test time in seconds (s) for the selected step.
Range: 0 or 0.3~999.0, 0 is for setting CONTINUE
Example: Input command SAFE:STEP2:AC:TIME 10
Description: This command sets the test time in STEP 2 to 10.0sec.
[:SOURce]:SAFEty:STEP<n>:AC:TIME[:TEST]?
This command queries the test time for the selected step.
Example: Input command SAFE:STEP2:AC:TIME?
Return message 1.000000E+01
Description: Return message 1.000000E+01 means the test time in STEP 2 is
10.0sec.
[:SOURce]:SAFEty:STEP<n>:AC:TIME:FALL < numeric value >
This command sets the time required for the voltage to fall to 0 in seconds (s) for
the selected step.
Range: 0 or 0.1~999.0 (0=OFF).
Example: Input command SAFE:STEP2:AC:TIME:FALL 3
Description: This command sets the time required for the voltage to fall to 0 for
STEP 2 to 3.0sec.
[:SOURce]:SAFEty:STEP<n>:AC:TIME:FALL?
This command queries the time required for the voltage to fall to 0 for the selected
step.

Table of Contents

Other manuals for Chroma 19032

Related product manuals