Electrical Safety Analyzer 19032 User’s Manual
5-22
[:SOURce]:SAFEty:STEP<n>:AC:LIMit:ARC:FILTer < numeric value >
This command sets the arc filter in Hertz (Hz) for the selected step.
Range : ARC bandwidth 23 kHz = 2.300000E+04
ARC bandwidth 50 kHz = 5.000000E+04
ARC bandwidth 100 kHz = 1.000000E+05
ARC bandwidth 230 kHz = 2.300000E+05
Example : Input command “SAFE:STEP2:AC:LIM:ARC:FILT 230000”
Description : This command sets the arc filter in STEP 2 to 230000Hz (230kHz).
[:SOURce]:SAFEty:STEP<n>:AC:LIMit:ARC:FILTer?
This command queries the arc filter for the selected step.
Example: Input command “SAFE:STEP2:AC:LIM:ARC:FILT?”
Return message “2.300000E+05”
Description: Return message “2.300000E+05” means the arc filter in STEP 2 is
230000Hz (230kHz).
[:SOURce]:SAFEty:STEP<n>:AC:TIME:RAMP < numeric value >
This command sets the ramp time in seconds (s) for the selected step.
Range : 0 or 0.1~999.0 (0=OFF).
Example : Input command “SAFE:STEP2:AC:TIME:RAMP 5”
Description: This command sets the ramp time in STEP 2 to 5.0sec.
[:SOURce]:SAFEty:STEP<n>:AC:TIME:RAMP?
This command queries the ramp time for the selected step.
Example: Input command “SAFE:STEP2:AC:TIME:RAMP?”
Return message “5.000000E+00”
Description: Return message “5.000000E+00” means the ramp time in STEP 2 is
5.0sec.
[:SOURce]:SAFEty:STEP<n>:AC:TIME[:TEST] < numeric value >
This command sets the test time in seconds (s) for the selected step.
Range: 0 or 0.3~999.0, 0 is for setting CONTINUE
Example: Input command “SAFE:STEP2:AC:TIME 10”
Description: This command sets the test time in STEP 2 to 10.0sec.
[:SOURce]:SAFEty:STEP<n>:AC:TIME[:TEST]?
This command queries the test time for the selected step.
Example: Input command “SAFE:STEP2:AC:TIME?”
Return message “1.000000E+01”
Description: Return message “1.000000E+01” means the test time in STEP 2 is
10.0sec.
[:SOURce]:SAFEty:STEP<n>:AC:TIME:FALL < numeric value >
This command sets the time required for the voltage to fall to 0 in seconds (s) for
the selected step.
Range: 0 or 0.1~999.0 (0=OFF).
Example: Input command “SAFE:STEP2:AC:TIME:FALL 3”
Description: This command sets the time required for the voltage to fall to 0 for
STEP 2 to 3.0sec.
[:SOURce]:SAFEty:STEP<n>:AC:TIME:FALL?
This command queries the time required for the voltage to fall to 0 for the selected
step.