Electrical Safety Analyzer 19032 User’s Manual
5-24
[:SOURce]:SAFEty:STEP<n>:DC:CURR:OFFSet?
This command queries the Offset value of the DC withstand test for the selected
step.
Example: Input command “SAFE: STEP2: DC: CURR: OFFS?”
Return message “+1.000000E-05”
Description: Return message “+1.000000E-05” means the Offset value of the DC
withstand test in Step 1 is set to 10uA.
[:SOURce]:SAFEty:STEP<n>:DC[:LEVel] < numeric value >
This command sets the DC voltage level in volts (V) for the selected step.
Range: 50~6000
Example: Input command “SAFE:STEP3:DC 4000”
Description: This command sets the DC voltage level for STEP 3 to 4000V.
[:SOURce]:SAFEty:STEP<n>:DC[:LEVel]?
This command queries the DC voltage level for the selected step.
Example: Input command “SAFE:STEP3:DC?”
Return message “4.000000E+03”
Description: Return message “4.000000E+03” means the DC voltage level for
STEP 3 is set to 4000V.
[:SOURce:]SAFEty:STEP<n>:DC:LIMit[:HIGH] < numeric value >
This command sets the DC withstand voltage leakage current high limit value in
amperes (A) for the selected step.
Range: 0.0000001~0.012
Example: Input command “SAFE:STEP3:DC:LIM 0.002999”
Description: This command sets the DC withstand voltage leakage current high
limit value in STEP 3 to 2.999mA.
[:SOURce:]SAFEty:STEP<n>:DC:LIMit[:HIGH]?
This command queries the DC withstand voltage leakage current high limit value
for the selected step.
Example: Input command “SAFE:STEP3:DC:LIM?”
Return message “2.999000E-03”
Description: Return message “2.999000E-03” means the DC withstand voltage
leakage current high limit value in STEP 3 is set to 2.999mA.
[:SOURce:]SAFEty:STEP<n>:DC:LIMit:LOW < numeric value >
This command sets the DC withstand voltage leakage current low
limit value in amperes (A) for the selected step.
Range: 0.0000001~0.012 (low limit value of leakage current ≤ high limit
value)
Example: Input command “SAFE:STEP3:DC:LIM:LOW 0.000001”
Description: This command sets the DC withstand voltage leakage current low limit
value in STEP 3 to 0.001mA.
[:SOURce:]SAFEty:STEP<n>:DC:LIMit:LOW?
This command queries the DC withstand voltage leakage current low limit value for
the selected step.
Example: Input command “SAFE:STEP3:DC:LIM:LOW?”
Return message “1.000000E-06”
Description: Return message “1.000000E-06” means the DC withstand voltage
leakage current in STEP 3 is set to 0.001mA.