GPIB/RS232 Operation Description (IEEE-488.2)
5-25
[:SOURce]:SAFEty:STEP<n>:DC:LIMit:ARC[:LEVel] < numeric value >
This command sets the arc current limit value in amperes (A) for the selected step.
Range: 0 or 0.0010~0.0300 (0=OFF.)
Example: Input command SAFE:STEP3:DC:LIM:ARC 0.0025
Description: This command sets the arc current limit in STEP 3 to 2.5mA.
[:SOURce]:SAFEty:STEP<n>:DC: LIMit:ARC[:LEVel]?
This command queries the arc current limit value for the selected step.
Example: Input command SAFE:STEP3:DC:LIM:ARC?
Return message 2.500000E-03
Description: Return message 2.500000E-03means the arc current limit in STEP
3 is set to 2.5mA.
[:SOURce]:SAFEty:STEP<n>:DC:LIMit:ARC:FILTer < numeric value >
This command sets the arc filter for the selected step.
Range: ARC bandwidth 23 kHz = 2.300000E+04
ARC bandwidth 50 kHz = 5.000000E+04
ARC bandwidth 100 kHz = 1.000000E+05
ARC bandwidth 230 kHz = 2.300000E+05
Example: Input command SAFE: STEP3: AC: LIM: ARC: FILT 230000
Description: This command sets the arc filter for STEP 3 to 230000Hz (230kHz).
[:SOURce]:SAFEty:STEP<n>:DC: LIMit:ARC:FILTer?
This command queries the arc filter for the selected step.
Example: Input command SAFE:STEP3:AC:LIM:ARC:FILT?
Return message 2.300000E+05
Description: Return message 2.300000E+05 means the arc filter for STEP 3 is
set to 230000Hz (230kHz).
[:SOURce]:SAFEty:STEP<n>:DC:TIME:DWELl < numeric value >
This command sets the dwell time in seconds (s) for the selected step.
Range: 0 or 0.1~999.0 (0=CONTINUOUS)
Example: Input command SAFE: STEP3: DC: TIME: DWEL 2.5
Description: This command sets the dwell time for STEP 3 to 2.5 sec.
[:SOURce]:SAFEty:STEP<n>:DC:TIME:DWELl?
This command queries the dwell time for the selected step.
Example: Input command SAFE: STEP3: DC: TIME: DWEL?
Return message 2.500000E+00
Description: Return message 2.500000E+00 means the dwell time for STEP 3 is
set to 2.5 sec.
[:SOURce]:SAFEty:STEP<n>:DC:TIME:RAMP < numeric value >
This command sets the ramp time step in seconds (s) for the selected.
Range: 0 or 0.1~999.0, 0 is for setting OFF
Example: Input command SAFE: STEP3: DC: TIME: RAMP 2
Description: This command sets the ramp time for STEP 3 to 2.0 sec.
[:SOURce]:SAFEty:STEP<n>:DC:TIME:RAMP?
This command queries the ramp time for the selected step.
Example: Input command SAFE: STEP3: DC: TIME: RAMP?
Return message 2.000000E+00