Electrical Safety Analyzer 19032 Users Manual
5-26
Description: Return message 2.000000E+00means the ramp time for STEP 3 is
set to 2.0 sec.
[:SOURce]:SAFEty:STEP<n>:DC:TIME[:TEST] < numeric value >
This command sets the test time in seconds (s) for the selected step.
Range: 0 or 0.1~999.0 (0=CONTINUOUS)
Example: Input command SAFE:STEP3:DC:TIME 1
Description: This command sets the test time for STEP 3 to 1.0sec.
[:SOURce]:SAFEty:STEP<n>:DC:TIME[:TEST]?
This command queries the test time for the selected step.
Example: Input command SAFE:STEP3:DC:TIME?
Return message 1.000000E+00
Description: Return message 1.000000E+00 means the test time for STEP 3 is
set to 1 sec.
[:SOURce]:SAFEty:STEP<n>:DC:TIME:FALL < numeric value >
This command sets the time required for the voltage to fall to 0 in seconds (s) for
the selected step.
Range: 0 or 0.1~999.0 (0=OFF)
Example: Input commandSAFE:STEP3:DC:TIME:FALL 3
Description: This command sets the time required for the voltage to fall to 0 for
STEP 3 to 3.0sec.
[:SOURce]:SAFEty:STEP<n>:DC:TIME:FALL?
This command queries the time required for the voltage to fall to 0 for the selected
step.
Example: Input command SAFE:STEP3:DC:TIME:FALL?
Return message 3.000000E+00
Description: Return message3.000000E+00means the time required for the
voltage to fall to 0 for STEP 3 is set to 3.0sec.
[:SOURce]:SAFEty:STEP<n>:DC:CHANnel[:HIGH] < channel list >
This command sets the high scan channels for the selected step. The <channel
list> format is: (@SN(C1, C2, C3)). SN is the scan box number; C1, C2, and C3 are
the channel numbers.
Example: Input command SAFE: STEP3: DC:CHAN(@2(1,2))
Description: This command sets the high scan channels for STEP 3 to BOX 2
channel 1 and 2 HIGH output.
Example: Input command SAFE: STEP3: DC:CHAN(@2(0))
Description: This command sets the high scan channels for STEP 3 to BOX 2 high
scan channel O F F.
[:SOURce]:SAFEty:STEP<n>:DC:CHANnel[:HIGH]?
This command queries the high scan channels for the selected step.
Example: Input command SAFE: STEP3: DC: CHAN?
Return message (@2(1,2))
Description: Return message (@2(1,2)) means the high scan channels for STEP
3 is set to BOX 2 channel 1 and 2 to HIGH output.
[:SOURce]:SAFEty:STEP<n>:DC:CHANnel:LOW < channel list >
This command sets the low scan channels for the selected step (RTN/LOW).
Example: Input command SAFE:STEP3:DC:CHAN:LOW (@2(2,4))