Specifications
19053, 19054: 15kg approx.
Safety
Ground Bond
Less than 100mΩ at 25Amp, 10sec
Hipot Less than 10mA at Wac 1.5kV, 3sec
Insulation
Resistance
Over 20MΩ at 500V 10sec
Line leakage current Less than 3.5mA at 127V, 3sec, normal, reverse
Note
1. AC set over 100 VA, DC set over 40VA the maximum operating time is 60
seconds, and the same as rest time. If the period is 1/2 duty (TUV ON), for
full rating output, the line input range is +10%, -0%.
2. Refer 1.2kV resistance load only. With the standard test lead, to get best
accuracy, please do not need to process OFFSET.
WAC mode is less than 500V add extra 3 counts error.
WAC scanners on, please add 10 counts/channel. WDC scanner on, add 2
counts/channel.
3. The best test time is dependent on device under test (DUT).
4. Validation point is 1.25kV with a 250kΩ resistor.
2-3