HIPOT Tester 19051/19052/19053/19054 User’s Manual
[: HIGH] <numeric_value>
: SOURce: SAFEty: STEP: AC: LIMit: HIGH
This command sets selected STEP which AC withstand voltage leakage
current high limit.
Example : Input command “SAFE:STEP 1:AC:LIM 0.01”
Description: This command sets AC withstand voltage leakage current
high limit value of the main unit STEP1 is 10mA.
[: HIGH]?
: SOURce: SAFEty: STEP: AC: LIMit: HIGH?
This command queries selected STEP which AC withstand voltage
leakage current high limit.
Example : Input command “SAFE:STEP 1:AC:LIM?”
Return message “1.000000E-02”
Description: Return message “1.000000E-02” means AC withstand
voltage leakage current high limit value of the main unit
STEP1 is set to 10mA.
: LOW <numeric_value>
: SOURce: SAFEty: STEP: AC: LIMit: LOW
This command sets selected STEP which AC withstand voltage leakage
current low limit.
Example : Input command “SAFE:STEP 1:AC:LIM:LOW 0.00001”
Description: This command sets AC withstand voltage leakage current
low limit value of the main unit STEP1 is 0.01mA.
: LOW?
: SOURce: SAFEty: STEP: AC: LIMit: LOW?
This command queries selected STEP which AC withstand voltage
leakage current low limit.
Example : Input command “SAFE:STEP 1:AC:LIM:LOW?”
Return message “1.000000E-05”
Description : Return message “1.000000E-05” means AC withstand
voltage leakage current low limit value of the main unit
STEP 1 is set to 0.01mA.
: ARC
[: LEVel] <numeric_value>
: SOURce: SAFEty: STEP: AC: LIMit: ARC: LEVel
This command sets selected STEP which ARC checking value.
Example : Input command “SAFE:STEP 1:AC:LIM:ARC 0.004”
Description: This command means ARC checking value of the main unit
STEP1 is set to 4mA.
[: LEVel]?
: SOURce: SAFEty: STEP: AC: LIMit: ARC: LEVel?
This command queries selected STEP which ARC checking value.
Example : Input command “SAFE:STEP 1:AC:LIM:ARC?”
Return message “4.000000E-03”
Description: Return message “4.000000E-03” means ARC checking
value of the main unit STEP 1 is set to 4.0mA.
: REAL
[: HIGH]
5-20